π§ͺ Soft particle glass flow curves, supplementary set B (Seth et al. 2011, Fig. S4b)ο
dataset id: seth2011_S4b
Steady-state flow curves of jammed soft-particle glasses digitized from supplementary Fig. S4b of Seth et al., Nature Mater. 10, 838-843 (2011). Three volume fractions (0.72, 0.75, 0.86). Companion to the S4a set, covering a second batch of the same soft-particle system; the paperβs micromechanical model is compared against both sets. One small digitization glitch in the 0.72 curve (single out-of-order point) was fixed by sorting.
π Dataο

ποΈ Metadataο
Field |
Value |
|---|---|
Material |
Soft particle glass (deformable microgel) |
Material kind |
microgel |
Source |
Seth, J. R.; Mohan, L.; Locatelli-Champagne, C.; Cloitre, M.; Bonnecaze, R. T.; A micromechanical model to predict the flow of soft particle glasses; Nature Materials; 2011 |
DOI |
|
Figure |
S4b |
Measurement type |
flow_curve |
Geometry |
cone-plate (50 mm, 2 deg (Anton Paar MCR 501)) |
Temperature |
20 Β°C |
Protocol |
Measured on an Anton-Paar MCR 501 with cone-and-Peltier-plate (50 mm, 2 deg, 48 um truncation) and solvent trap at 20 C; roughened shearing surfaces (waterproof sandpaper, 20 um) to suppress wall slip. |
Provenance |
literature, digitized: yes |
π§« Samplesο
Sample id |
Label |
|---|---|
phi_0.86 |
Ο = 0.86 |
phi_0.75 |
Ο = 0.75 |
phi_0.72 |
Ο = 0.72 |
π Data columnsο
Role |
Column |
Units |
|---|---|---|
|
|
1/s |
|
|
Pa |
|
|
β |